+63 2 4649339

Gas pressures from 100 kPa to 110 MPa (15 to 16,000 psi)

PG7202 is a member of the PG7000 family of reference level piston gauges that covers the range of high pressure gas up to 110 MPa (16,000 psi). Operating directly in gas avoids difficult to use liquid/gas separators that add uncertainty to the measurement process. PG7202’s gas operated, liquid lubricated piston-cylinder design overcomes the disadvantages of lubricating the piston-cylinder with gas at high pressure. Performance is not affected by gas cleanliness and piston drop rates are even lower than those typically found in hydraulic piston gauges.

PG7202 is combined with the GPC1 gas pressure controller to configure a complete manually operated calibration system with effort free gas pressure setting and adjusting. By substituting a PPCH-G automated pressure controller for GPC1 and adding on AMH-100 automated mass handler, PG7202 operation can be fully automated.

All PG7202 piston-cylinder modules can be delivered using Krytox®, a fluorinated synthetic oil, to lubricate the piston-cylinder for applications in which the system must be free of hydrocarbons.

Differential pressure at elevated line pressure

Two PG7202s can be used in combination to define low differential pressures at elevated line pressures. Contact Fluke Calibration for more information.

General

Power requirements 85 V ac to 264 V ac, 47 Hz to 440 Hz, 22 VA max consumption
Normal operating temperature range 18 °C to 28 °C (64 °F to 82 °F)
Weight Instrument platform with no mass loaded: 13 kg (28 lb)
PG terminal 1.4 kg (3 lb)
Dimensions Instrument platform (HxWxD): 36 cm x 40 cm x 35 cm (14.5 in x 15.8 in x 13.8 in)
Height: Top of mounting post with piston-cylinder module installed
PG terminal (HxWxD): 12 cm x 15 cm x 20 cm (4.7 in x 5.9 in x 7.9 in)
Communications ports RS-232 COM1: Host computer
COM2: AMH automated mass handler, external barometer
COM3: Automated pressure controller
IEEE-488.2: Host computer
Maximum pressure ranges Actual range depends on piston-cylinder and mass set selection.
Operating media Gas Operation: air, nitrogen
Oil Operation: ethyl hexyl Sebacate (oil)
Maximum mass load 100 kg
Pressure connections Test port: DH500Note: DH200 and DH500 are gland and collar type fittings for 1/4 in (6 mm) coned and left hand threaded tube. DH200 is equivalent to AE SF250C, HIP LF4, etc. DH500 is equivalent to AE F250C, HIP HF4, etc.
CE conformance Available, must be specified