9118A Thermocouple Calibration Furnace

CONTACT US NOW FOR QUOTATION!!

+632 82515165 / +632 82570795

High performance furnace for thermocouple calibrations to 1200 °C

The Fluke Calibration 9118A Thermocouple Calibration Furnace is a horizontal, open-ended tube furnace with a temperature range of 300 °C to 1200 °C. It is used for comparison calibration of noble and base-metal thermocouples by secondary high-temperature labs and instrument shops in industries such as aerospace, automotive, energy, metals, and plastics. The 9118A is the most accurate, reliable, and flexible furnace in its class, meeting the demanding requirements of high-temperature thermocouple calibration.

General specifications

Operating conditions
Operating temperature 5 °C to 40 °C
Storage temperature –20 °C to 70 °C
Humidity 80 % maximum for temperatures <31 °C, decreasing linearly to 50 % at 40 °C
Altitude <2,000 m (6,562 feet)
Power requirements 230 V ac (±10 %), 50/60 Hz, 20 A
Heater power 4000 watts at 230 V ac
Over-current protection
System 20 A, 250 V resettable circuit breaker
Main heater fuse F 12 A, 250 V
Zone heater fuse F 12 A, 250 V
Computer interface RS-232 and USB
Display Monochrome LCD, °C or °F user-selectable
Display resolution 0.1 °C or °F
Size (H x W x L) 400 mm x 337 mm x 700 mm (15.7 in x 13.3 in x 27.6 in)
Net weight 29 kg (63.9 lb) without isothermal block
Isothermal block (optional)
Isothermal block construction Alumina
Block outer diameter 37 mm (1.5 in)
Block length 380 mm (15.0 in)
Well diameter (four wells total) 6.7 mm (0.26 in)
Well depth 365 mm (14.4 in)
Net weight 0.84 kg (1.9 lb) isothermal block only

Accuracy specifications

All accuracy specifications except temperature are for a period of 1 year after calibration, at 13 °C to 33 °C
The furnace can be used with or without an Isothermal Block installed. Unique calibration parameters are necessary for each configuration. Calibration does NOT come standard from the factory for each configuration. Ensure that the furnace has been properly calibrated for the desired mode of operation.
Furnace temperature range 300 °C to 1200 °C
Set-point accuracy ± 5 °C

 

Accuracy specifications

All accuracy specifications except temperature are for a period of 1 year after calibration, at 13 °C to 33 °C
The furnace can be used with or without an Isothermal Block installed. Unique calibration parameters are necessary for each configuration. Calibration does NOT come standard from the factory for each configuration. Ensure that the furnace has been properly calibrated for the desired mode of operation.
Furnace temperature range 300 °C to 1200 °C
Set-point accuracy ± 5 °C