High performance furnace for thermocouple calibrations to 1200 ?C
The Fluke Calibration 9118A Thermocouple Calibration Furnace is a horizontal, open-ended tube furnace with a temperature range of 300 ?C to 1200 ?C. It is used for comparison calibration of noble and base-metal thermocouples by secondary high-temperature labs and instrument shops in industries such as aerospace, automotive, energy, metals, and plastics. The 9118A is the most accurate, reliable, and flexible furnace in its class, meeting the demanding requirements of high-temperature thermocouple calibration.
General specifications |
|
Operating conditions | |
Operating temperature | 5 ?C to 40 ?C |
Storage temperature | ?20 ?C to 70 ?C |
Humidity | 80 % maximum for temperatures <31 ?C, decreasing linearly to 50 % at 40 ?C |
Altitude | <2,000 m (6,562 feet) |
Power requirements | 230 V ac (?10 %), 50/60 Hz, 20 A |
Heater power | 4000 watts at 230 V ac |
Over-current protection | |
System | 20 A, 250 V resettable circuit breaker |
Main heater fuse | F 12 A, 250 V |
Zone heater fuse | F 12 A, 250 V |
Computer interface | RS-232 and USB |
Display | Monochrome LCD, ?C or ?F user-selectable |
Display resolution | 0.1 ?C or ?F |
Size (H x W x L) | 400 mm x 337 mm x 700 mm (15.7 in x 13.3 in x 27.6 in) |
Net weight | 29 kg (63.9 lb) without isothermal block |
Isothermal block (optional) | |
Isothermal block construction | Alumina |
Block outer diameter | 37 mm (1.5 in) |
Block length | 380 mm (15.0 in) |
Well diameter | (four wells total) 6.7 mm (0.26 in) |
Well depth | 365 mm (14.4 in) |
Net weight | 0.84 kg (1.9 lb) isothermal block only |
Accuracy specifications |
|
All accuracy specifications except temperature are for a period of 1 year after calibration, at 13 ?C to 33??C | |
The furnace can be used with or without an Isothermal Block installed. Unique calibration parameters are necessary for each configuration. Calibration does NOT come standard from the factory for each configuration. Ensure that the furnace has been properly calibrated for the desired mode of operation. | |
Furnace temperature range | 300 ?C to 1200 ?C |
Set-point accuracy | ? 5 ?C |
Accuracy specifications |
|
All accuracy specifications except temperature are for a period of 1 year after calibration, at 13 ?C to 33??C | |
The furnace can be used with or without an Isothermal Block installed. Unique calibration parameters are necessary for each configuration. Calibration does NOT come standard from the factory for each configuration. Ensure that the furnace has been properly calibrated for the desired mode of operation. | |
Furnace temperature range | 300 ?C to 1200 ?C |
Set-point accuracy | ? 5 ?C |